Protein detection based on microelectrodes with the PPy[3,3-Co(1,2-C2B9H11)]2 solid contact and immobilized proteinases: preliminary investigations

Мова статті: 
Англійська
Автор(и) статті: 
O.A.Biloivan, S.V.Verevka, S.V.Dzyadevych, N.Jaffrezic-Renault, N.Zine, J.Bausells, J.Samitier, A.Errachid
Вихідні данні статті: 
Material Science and Engineering C.- Volume 26, Issue 2-3, March 2006, Pages 574-577
Назва журналу: 
Material Science and Engineering C
Рік публікації: 
2006
Volume: 
26
Issue: 
2-3
Ключові слова: 
Immobilized proteinase; Microbiosensor; Protein; Solid-contact ion-selective microelectrodes (SCISME)
Ключове слово: 
Immobilized proteinase
Ключове слово: 
Microbiosensor
Ключове слово: 
Protein
Ключове слово: 
Solid-contact ion-selective microelectrodes
Ключове слово: 
SCISME
Protein selective potentiometric microelectrodes with the PPy[3,3-Co(1,2-C2B9H11)]2 solid internal contact were fabricated by the immobilization of different proteinases such as trypsin, pronase E and carboxypeptidase B on the surface of platinum microelectrodes modified by a conducting polymer. The performance of these biosensors was investigated by using potentiometric measurements. The feasibility of this biosensor for protein determination was demonstrated, with a detection limit of about 1-2 μg/ml for protamine in buffer solution. © 2005 Elsevier B.V. All rights reserved.